I
NITIAL
S
TARTUP AND
C
ALIBRATION
3-12
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
and will move cursor right/left, one character at a time.
and will return analyzer to RUN MODE
SHIFT
ENTER
±
5%
Must be between 700mA and 950mA
700mA to 800mA for % to high ppm ranges
800mA to 950mA for mid to low ppm.
Adjust R9 on Power Supply Board (See Figure 3-1).
Refer to Data Sheet for initial factory setting.
75 to 80% of maximum.
De-tuned in counterclockwise direction.
0.4V to 1.2V with zero gas flowing (see Sections 5.6 and 5.7)
1.2V to 7.5V with span gas (see Section 5.4)
7.5V max. On highest range.
Adjust R3 on Signal Board (see Figure 3-1).
ENTER
SHIFT
DIAGNOSTICS
SHIFT
SRC.CUR.=XXXmA
SHIFT
OSC. T =
DET. SIG. - X.XX
+5V = X.X
SHIFT
±
5%
-15V = XX.X
SHIFT
SHIFT
+15V = XX.X
±
5%
±
5%
+12V = XX.X
2. Connect span gas to the sample cell inlet at the back of the analyzer. Flow the gas
at a rate of 500 to 1000 cc/min (1 to 2 SCFH), as read on a flowmeter, until the
display reading stabilizes.
3. To calibrate, press SPAN and then ENTER.
F
IGURE
3-7. D
IAGNOSTIC
M
ODE
3.10 LINEARIZATION
The Model 880A Analyzer can be operated in the linear and non-linear mode.
Linearization can be toggled ON/OFF with the
↑
or
↓
key in the Linearization Mode
Screen (see Figure 3-8) In the OFF position, linearization is disabled for all ranges. In
the linear mode, the component of interest is measured in engineering units, either
ppm (parts per million) or % (percent of composition). In the non-linear mode, the
component of interest is measured is in %FS (percent of fullscale).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 880A
Страница 83: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 85: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 87: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...