I
NITIAL
S
TARTUP AND
C
ALIBRATION
3-10
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
Note:
The instrument will not allow the user to enter a value in the [CAL GAS = XXXX]
screen that is larger than the fullscale value entered in the [FS = XXXX ppm)
screen.
3.6.5 C
ALIBRATION
G
AS
In the [CAL GAS=XXXX ppm *] display, up to a four-digit calibration gas value is
entered in ppm for each linearized range.
It is recommended that cal gas
concentration be from 80% to 100% of fullscale.
3.6.6 Z
ERO
O
FFSET
In the [ZR-OFFSET:X.XX *] display, the amount of zero offset in percent of fullscale is
entered for each range. The zero offset feature compensates for impurities in zero
calibration gas. If there are no impurities in the zero gas, set ZR-OFFSET to 0.00.
3.6.7 T
IME
C
ONSTANT
In the [TIME CONST=XX
←
display, the value of the TIME CONSTANT can be
changed for each range. This TIME CONSTANT is responsible for the amount of time
(in seconds) in which the analyzer responds to change. A different TIME CONSTANT
can be selected for each range.
To change or check the settings of the different range parameters, follow the steps in
the appropriate figure in this section.
Note:
After changing a setting, press ENTER to retain the new setting in nonvolatile
memory. Settings stored in nonvolatile memory will be saved even in the event
of a power outage.
At any point in the sequence, the routine may be exited by pressing
SHIFT/ENTER.
The analyzer must be in LOCAL mode (L1, L2 or L3 appears in RUN MODE display)
to change ranges in [RANGE=XXX] screen of Figure 3-6.
3.7 ANALYZER DIAGNOSTICS
Diagnostics are selectable through the MODE function. This function allows the
source current, detector signals, oscillator tune voltage, and the four power supply
values (+5 V, +15 V, +12 V, and -15 V) to be displayed.
It is recommended that the values for oscillator tune, detector signal and source
current be recorded when the diagnostic display is first accessed. See Figure 3-7.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 880A
Страница 83: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 85: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 87: ... blank Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...