3-10
MODEL 54 C
SECTION 3
PROGRAMING THE ANALYZER
5.
Pressing More (F3) allows the configuration of timer setting.
*Delays relay deactivation after reading is no longer in the alarm 3 condition
6.
NOTE
The timer can be used for periodic chemical or mechanical
cleaning of a coated sensor. See the
cycle below. The timer must be activated separately (see
Pressing More (F3) allows configuration of more timer parameters.
7.
Pressing Exit (F1) returns screen to previous functions.
TIMER EXAMPLES
A3:
High
A3 Hys:
000 %
A3 Delay:
0 Sec
Exit
More
Enter
Timer setting:
Interval:
000.0 hr
Repeats:
01
Exit
More
Enter
More timer settings
On:
1
20 Off: 1 sec
Recovery: 600 sec
Exit
More
Enter
Menu Item
Options
A3 alarm 3: Configuration
Low alarm/High alarm/
Off (not used)
A3 alarm 3: Hysteresis (deadband)*
0-25 %
A3 alarm 3: Delay time before relay activation
and deactivation (can prevent chatter)
0-99 sec
Menu Item
Options
Interval between cycles
0-999.9 hr.
No. of activation cycles for relay (repeats)
1-60
Menu Item
Options
Relay “on” (time per activation)
0-999 sec
Relay “off” (time between activations)
0-999 sec
Recovery time after last activation
0-999 sec
RELAY
ACTIVATION
TIME
Interval
(hr)
Recovery
(sec)
“On”
Time
(sec)
One activation cycle
RELAY
ACTIVATION
TIME
Interval
(hr)
Recovery
(sec)
“On”
Time
(sec)
“Off”
Time
(sec)
Two activation cycles (repeats)