3-6
MODEL 54 C
SECTION 3.0
PROGRAMING THE ANALYZER
This section describes the procedure for changing what is shown on the
operating display.
1.
From the program menu, move the cursor to “Configure” and press
Enter (F4). The first configuration menu is displayed. With the cur-
sor on “Display” press Enter (F4). Pressing More (F3) allows the
configuration of other display options.
2.
Press Enter (F4) to accept the value shown in reverse video and
move the cursor to the next selection. Press More (F3) to show other
display options.
3.
Press Enter(F4) to accept the option displayed and move the cursor
to the next option. Press More (F3) to select other menu items.
4.
This screen allows you to choose the items displayed on the third
line right and left of the operating display screen. The process tem-
perature and output 1 value are always shown on line 2 of the oper-
ating display. This screen allows you to make the following choices:
The “Display contrast” selection makes the display become lighter or
darker. 0 is the lightest display and 9 is the darkest. The display is
immediately updated. Press Exit (F1) to return to the first configura-
tion menu.
3.3 CONFIGURATION OF DISPLAY OPTIONS
Lower Left of Operating Display
• AL 1 (alarm 1 setpoint)
• AL 3 (alarm 3 setpoint)
• ABS (absolute conductivity)
• CF (cell factor)
• Blank (nothing displayed on lower left of
operating display)
Lower Right of Operating Display
• AL 2 (alarm 2 setpoint)
• AL 3 (alarm 3 setpoint)
• ABS (absolute conductivity)
• Slp (temperature slope)
• Out 2 (Output 2 value in mA or %)
• Blank (nothing displayed on lower right of
operating display)
Display
Outputs
Alarms
Exit
More
Enter
Sensor:
Inductive
Ref temp: 25.0°C
Temp units:
°C
Exit
More
Enter
Output 1 units:
%
Output 2 units: mA
Language: English
Exit
More
Enter
Display left:
AL1
Display right:
AL2
Display contrast: 4
Exit
More
Enter
Menu Item
Options
Sensor type
Inductive/Contacting
Reference temperature
0-200°C
Temperature units
°C/°F
Menu Item
Options
Output 1 units
mA/percent
Output 2 units
mA/percent
Language
English/Français/Español/Deutsch/Italiano