Connecting the Lead Wires to the TOP TENS™
Hold the insulated portion of the electrode connector and push the plug end of the lead wire into the
electrode connector. Repeat for all 4 electrodes. After connecting each electrode, plug the other end
of the lead wire into the TOP TENS™.
Caution: The TOP TENS™ should be OFF before connecting the lead wire to the TOP TENS™.
Note:
Use care when you plug and unplug the wires. Pulling on the lead wire instead of its
insulated connector may cause wire breakage.
Caution
: Never insert the plug of the lead wire into an AC power supply or other equipment not
specified as safe for the lead wires.
Turning on the TOP TENS™
•
Turn on TOP TENS™ device by pressing the On/+ button.
Setting Treatment Program, Time and Intensity Level
•
Press the Mode button once to select treatment program, default P1 will flash on the screen
-
Press the On/+ (increase) or Off/- (decrease) button to scroll through treatment programs
P1 thru P8 (see page 14 for treatment program details).
-
Press the Mode button to lock in the selected treatment program and to open up treatment
time selection. Treatment program will stop blinking on the screen, and the default treatment
time of 30 minutes will blink on the screen.
-
While treatment time is blinking press the On/+ (increase) or Off/- (decrease) button
to scroll through treatment time in 5 minute increments from 5 to 60 minutes and C
(continuous). Once you reach 60 and press the On/+ button one more time, C (continuous
time) will appear.
-
Press mode button to lock in treatment time.
•
Press the CH button to set the Channel (lead wire connection) Intensity. Channel 1 (left side of
screen) is the set default and will be blinking.
•
Press the On/+ (increase) or Off/- (decrease) button to adjust stimulation intensity for channel 1.
•
Press the CH button again to set Channel 2 (lead wire connection) Intensity. Channel 2 (right side of