Low Battery Indicator
When the low power indicator
"
"
displays on the LCD, stop the
device and charge the battery.
Charging the Battery
NOTE:
Battery comes pre-charged.
Proceed as follows to recharge the battery:
• This device cannot be used while charging.
• Make sure that the device has been switched off.
• Make sure that the device is no longer connected to
the patient (the output cables and electrodes must be
disconnected).
• Connect the USB cable to the charging port on the top of device.
• Connect the USB cable to the wall outlet or USB port.
• When the device is charging, the indicator light will be red.
• It could take up to 6 hours to reach a full charge.
• When charging is completed, the indicator light will be green.
After the battery has been recharged, disconnect the cable from the wall
outlet or USB port and the device is ready to be used again.
The life of a rechargeable battery depends on the number of recharging/
rundown cycles it undergoes and how these cycles are performed.
The following suggestions will help prolong the life of the battery:
• Whenever the device is not used frequently, charge
the battery once a month.
• For longer battery life, discharge the battery as much
as possible.
CAUTION
1) Please use the standard charger provided by the manufacturer or agent.
2)
Do not use the device for treatment when in charging status as
it will not work. When you are using the device, never connect it
to the charger. If you do, the device will turn off automatically as
a safety precaution and you will not be able to turn it back on until
you disconnect the charger from the device.
3) When charging is complete, you are strongly advised to disconnect
the charger