![Roscoe Medical BUNNY Series Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/roscoe-medical/bunny-series/bunny-series_manual_1486416013.webp)
13
Figure 12
Air Filter Cover
Air Filter
Kit: Mouthpiece, TruNeb Reusable Nebulizer Cup & 7' Tubing
50016
Air Filters (6)
CIF-1128PK6
Nylon Carry Bag
BAG-BUNNY
AIR FILTER:
• The air filter on the side panel of the compressor unit is bright white when
new. With use, it will turn gray, which means it is time to replace it. It must
be replaced immediately if it gets wet or becomes clogged.
• To replace the air filter, ensure the compressor is turned off.
FILTER CHANGE
1.
DO NOT
wash or clean the filter.
ONLY
use filters supplied by your
distributor and
DO NOT
operate without a filter.
2. Check the filter every 15 days. Change the filter when it is visibly dirty
OR
every 30 days, whichever comes first.
3. Changing procedure (Fig. 12):
a. Remove the filter cover by twisting counter-clockwise and pulling out.
b. Replace the used filter with a new one.
c. Replace the filter cover.
CAUTION
DO NOT
attempt to wash and reuse the air filter.
DO NOT
use substitute
filter material. Compressor damage may result.