4.6 Wire diagram
The changing bed is delivered with the control box already mounted on the frame. The diagram
shows how the cables are connected to the control box.
All the plugs are watertight, and therefore they can be a
little tricky to connect / disconnect. Make sure that the
plugs are positioned correctly before they are pushed
into the sockets at the control box. The plugs must be
pushed into the sockets until the “O-ring” is no longer
visible.
Motor, control box and control switch apply to IPX6. This
means that they are dustproof and water resistant.
The electronic system is designed for periodical use
only, this means a max. duty cycle of 10%; 2min on /
18min off.
1. Control box
2. Mains cable 230V
3. Motor, connects to socket 1
4. End plugs, connects to socket 2, 3 & 4
5. Handcontrol, connects to HB socket