![Ropox Hudson 40-25023 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/ropox/hudson-40-25023/hudson-40-25023_user-manual-mounting-instruction_1486182020.webp)
Page 20
Electrical components
Connecting cables
The shower bed is delivered with the control box already mounted on the frame.
Motor cable connects to socket 1. Socket 2, 3 & 4 are not in use and are blocked.
Control switch connects to ”HB”.
When all cables have been connected, use the safe
locking bracket. Pushed in until you hear a “click”.
Mains cable connects to the socket at the side of the
box.
Disconnecting cables
Mains cable can be pulled out (no safe locking bracket).
Before the other cables can be disconnected, the safe locking bracket needs to be
released / removed (
use a screwdriver to remove
). Locking clips at each end of the safe
locking bracket have to be released before the bracket can be removed, and the cables
can be disconnected.
Placement of the handcontrol.
Bracket for hand control can be placed at the wall
as shown. To be mounted with the two screws
which are delivered with the bracket.
To obtain the full IP66 value, it is very important that all the plugs have been pressed
firmly into the sockets and the safe locking bracket is mounted.
Hand control should be placed out of the reach of children