16
TF 200.02.0012_eng
6.
Maintenance
Note!
Failing to comply with periodic maintenance may result of degrading of product function and
safety.
Notification!
Ropox will make parts list and circuit diagrams available for maintenance and repair beyond
what is described in this manual. Repair beyond what is described in this manual may only be
done by Ropox designated service personnel.
6.1 Periodic maintenance
Once a year:
Adjust the bathtub to top position and check that:
1. All bolts on the top and bottom of the actuator are securely tightened.
2. All connections (water supply and drain) are in order and undamaged.
3. Mixer tap, shower hose and shower head are in order and undamaged.
4. The frame and cover plates are undamaged.
5. Press the DOWN button, adjust the bathtub to bottom position and check that the movement is
smooth and even. See that the hose connections follow the movement of the bathtub without getting
jammed.
6. Now press the UP button, adjust the bathtub to top position, and check that the movement is smooth
and even. See that the hose connections follow the movement of the bathtub without getting
jammed.
7. If a safety strip has been mounted. Press the DOWN button, let the bathtub move 2-5 cm downward,
and activate the safety strip with a light pressure. The bathtub should now stop the downward
movement, move 1-2 cm upward and stop. If the safety strip does not function, please contact
Ropox or Ropox representative.
6.2 Trouble shooting
Problem
Remedy
The bathtub does not move up
or down when the hand-control
switch is pressed
Check whether:
1. 230V has been connected and the outlet is working.
2. The motors are connected with the control unit.
3. The handset has been connected with the control unit .
4. The motors are overloaded (note: 2 min. ON, 18 min
OFF)
Control switch defective
Check whether:
1. The bathtub works with another control switch
When pressing the UP and
DOWN buttons you only hear
“clicks” from the control unit,
bathtub does not move
1. The plugs may not have been inserted correctly.
Dismount all cables from the control unit and insert them
again.
Resetting may be required.
The bathtub is not horizontal /
level
”Reset”
Press and hold UP ( ) and DOWN ( ) buttons. Keep the
buttons depressed until the control unit has beeped approx. ten
times. After that, the bathtub may be adjusted to the bottom
position for resetting.