Dimensions
Version 1
RES-401
Page 29
9
Dimensions
10
Modifications (MODs)
Modifications are not available for the RES-401.
Installation
If several controllers are installed on one top hat
rail (DIN TS35 rail), a clearance of at least 20mm
should be allowed between them.
The moving clip required for fastening must be
facing down for mounting on a horizontal top hat
rail.
End holders to mechanical fix the controller must be fitted at both ends for
mounting on a vertical top hat rail.
Weight
Approx. 0.7kg (incl. connector plug-in parts)
Housing material
Plastic, polycarbonate, UL-94-V0
Connecting cables
Type / cross-sections
Rigid or flexible; 0.2…2.5mm² (AWG 24…12)
Plug-in
If ferrules are used, they must be crimped in accordance with
DIN 46228 and IEC/EN 60947-1.
This is essential for proper electrical contact in the terminals.
!
75.0
90.0
11
3.
0
13
5.
0