Preparing for Use
R&S
®
NRX
15
Getting Started 1424.7070.02 ─ 05
NRQ6
Frequency Selective Power Sensor
LAN PoE+
Host
Interface
Sample Clock I/O (Default: 120 MHz)
LO I/0
Figure 4-1: Setup with an R&S power sensor (example)
1 = Signal source
2 = R&S power sensor
3 = Host Interface connector
4 = R&S NRP
‑
ZK8
5 = Sensor connector of the R&S NRX
6 = R&S NRX
Use an R&S NRP
‑
ZK8 cable to connect an R&S power sensor to the R&S NRX.
If you use an R&S NRP
‑
ZK6 cable, the reference clock and trigger are not sup-
ported.
1. 8
-
pin female connector of R&S NRP
‑
ZK8:
a) Insert the screw-lock cable connector into the host interface of the R&S
power sensor.
b) Tighten the union nut manually.
2. 8
-
pin male connector of R&S NRP
‑
ZK8:
a) Insert this connector into one of the sensor ports of the R&S NRX.
3. Connect the RF connector of the R&S power sensor to the signal source. For
details, see the user manual of the R&S power sensor.
Note:
Incorrectly connecting/disconnecting an R&S power sensor can dam-
age the power sensor or lead to erroneous results.
4.8.2
Optional Sensor Interface for R&S NRT (R&S NRX-B9)
See
Suitable for R&S
NRT directional power sensors.
► Connect the R&S NRT-Zxx power sensor between source and load.
Connecting Power Sensors