23
(IEC 60601-1-2)
Table 6: Recommended separation distances between portable and mobile RF
communications equipment and not life supporting equipment
The
RomiApex™ A-15 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
RomiApex™ A-15 can help prevent
electromagnet interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
RomiApex™ A-15 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 0.35
P
80 MHz to 800 MHz
d
= 0.35
P
800 MHz to 2.5 GHz
d
= 0.7
P
0,01
0,04
0,04
0,07
0,1
0,11
0,11
0,22
1
0,35
0,35
0,7
10
1,1
1,1
2,2
100
3,5
3,5
7
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the higher frequency range applies.
Note 1: At 80 MHz and 800 MHz, the separation distance fort the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.