15
#
Operation
Instructions
Details and Warnings
- Drying time after
sterilization
– 30
minutes.
- The only sterilization
parameters to be used are
those that have been provided
in this Manual.
1
0
Storage
Keep devices in
sterilization packaging
in a dry and clean
environment.
Sterility cannot be guaranteed if
packaging is open or damaged
(check the packaging before
using the instruments).
8. Troubleshooting Guide
Problem
Possible Solution
Blank screen after use
- The device automatically shuts down after 5 minutes
without use - press the On-Off button to turn the device
on.
- Battery is completely dead
– replace with fresh battery -
ensure the battery is installed with proper polarity. Realign
if necessary.
- Device has malfunctioned
– contact your supplier.
Display showing no
progression of the file
towards the apical
area/foramen
- If already in use, this may indicate a poor connection -
check all connections and ensure lip clip is contacting the
oral mucosa and file clip is clean and free of debris
- Fill the canal with an electrolytic solution, if required (i.e.
Sodium Hypochlorite, Saline, etc.).
- The endodontic file is not touching the internal canal walls
- replace the file using a larger diameter file.
- If the behavior persists, the measuring cable or file clip
may need to be replaced and/or the device should be sent
in for service
– contact your supplier.
The cable icon
does not appear
- Make sure the measuring cable is properly connected.
- If the behavior persists, the measuring cable may need to
be replaced and/or the device should be sent in for service
– contact your supplier.
The File icon
keeps blinking
- The file clip is not properly connected with the endodontic
file.
- Check all connections and ensure lip clip is contacting the
oral mucosa and file clip is clean and free of debris.
- If the behavior persists, the measuring cable may need to
be replaced and/or the device should be sent in for service
– contact your supplier.
The connection icon
does not appear
- Try connecting another file clip to the measuring cable.