33
ranging from 1 (unathletic) to 4 (athletic)
(Fig. 14).
Choose the solid cup according to the
size of the distal end of the stump. (Fig. 15)
The stump must not come in contact
with the rim of the solid cup (danger of
deformation). If a solid cup that is too lar-
ge is chosen, there is too much space
between the stump and membrane.
Choose the size of the distal stump end
bearing cushion that fits the solid cup.
Recheck your choice for the patient.
Next, determine the correct height in
the system and then measure the length
from the distal end of the stump to the
desired socket height. Calculate the dif-
ference in regard to the entire length of
the height-compensating discs, always
beginning with the distal end cap.
(Fig. 16)
(500mm – assumed socket length = sys-
tem length to be set)
Screw the distal end cap, the height-
compensating discs and the solid cup
(including foam cushion) together
(Fig. 17).
Empty the water completely from the
cylinder, using the reset button (Fig. 18).
Carefully position the spacer assem-
bly you have screwd together into the
Fig. 16
Fig. 17
Fig. 18
Fig. 14
Fig. 15