Maintenance
The lift must undergo thorough inspection at least once per year. Inspection must be performed by authorized
personnel and in accordance with Direct Healthcare Group’s service manual.
Repairs and maintenance may only be done by authorized personnel using original spare parts.
Used batteries are to be left at the nearest recycling station. Used batteries can also be returned to
Direct Healthcare Group or a Direct Healthcare Group dealer for recycling.
Service agreements
Direct Healthcare Group offers the possibility of service agreements for maintenance and regular testing of your mobile
lift. Contact your local Direct Healthcare Group representative.
Cleaning/disinfection
If necessary, clean the lift with a cloth with warm water or a soap solution and check that the castors are free from dirt
and hair. The electronic components (battery, control box, hand control, actuators, cables) should be cleaned with a
damp cloth only. To avoid degreasing of the piston rods, the actuators should be retracted to minimum stroke and
without load before cleaning. Ensure that the lift is dried thoroughly after cleaning. Do not steam clean due to risk of
corrosion. Do not use cleaning agents containing phenol or chlorine, as this could damage the materials. If disinfection is
needed, 70 % ethanol, 45% isopropanol or similar should be used. Recommended frequency of cleaning is weekly and
possibly more often depending on frequency of use.
Storage and transportation
If the lift is not to be used for some time or e.g., during transport, we recommend that the emergency stop button be
pressed in. The lift should be transported and stored in -10 ° C to + 50 ° C and in normal humidity, 20% to 80% non-
condensing. The air pressure should be between 700 and 1060 hPa. See also Technical Information below, and marked
on the device. Leftmost symbol indicates storage and transportation. Let the lift reach room temperature before the
batteries are charged or the lift is used. The lift should not be stored so that it is exposed to dust, or so that the battery is
exposed to direct sunlight.
Operation
The operating environment should be 5 °C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa. See also Technical Information below, and marked on the device. Leftmost symbol indicates
operating condition.
40°C
5°C
50°C
-10°C
1060
hPa
700
hPa
1060
hPa
700
hPa
%
80%
20%
%
80%
20%
40°C
5°C
50°C
-10°C
1060
hPa
700
hPa
1060
hPa
700
hPa
%
80%
20%
%
80%
20%
15
I F U
System
Содержание Vega505EE
Страница 19: ...19 I F U System ...
Страница 36: ...36 I F U System ...
Страница 41: ...Bruke produktet 41 I F U System ...
Страница 53: ...53 I F U System ...
Страница 70: ...70 I F U System ...
Страница 75: ...Tuotteen käyttö 75 I F U System ...
Страница 87: ...87 I F U System ...
Страница 104: ...104 I F U System ...
Страница 109: ...Het product gebruiken 109 I F U System ...
Страница 121: ...121 I F U System ...
Страница 138: ...138 I F U System ...
Страница 155: ...155 I F U System ...
Страница 172: ...172 I F U System ...
Страница 176: ...176 I F U System ...