• MiniLift125must be assembled in accordance with the assembly instructions provided by RoMedic.
• Warranty does not apply if repairs or alterations are made by personnel who are not authorized by RoMedic.
• Batteries, control unit and actuators on MiniLift125 must not come in direct contact with water.
• To ensure optimal function, the lift must be inspected regularly. See section on other maintenance details.
• Never move the lift by pulling on the actuator!
• The lift may only be used on a level floor.
• Under no circumstances may max. load be exceeded. See section on technical information and the product decal on
the lift.
• Lifting accessories must be properly trial fitted and tested in relation to the patient’s needs and functional ability.
• MiniLift125 must not be charged in a wet room or used in connection with showering and bathing.
• The lift must not be exposed to running water.
Do not push
Technical information
Lifting speed:
3.5 cm/s (1 1/2 inch/s) without load
Batteries:
Two 12V 2.7 Ah valve-regulated, sealed, lead accumulator (gel-type batteries)
Charger:
Max. 400mA
Motor (mast):
DC 24 V, 5 A. IP X4. Operating time: 10% maximum continuous in 2 minutes, max 5 directional
changes per minute. Push: 10,000N.
Motor (base):
DC 24 V, 5 A IP X4. Operating time: 10% maximum continuous in 2 minutes, max 5 directional
changes per minute. Push: 1,000N.
Material:
Steel
Emergency lowering:
Mechanical
Wheels:
Rear 100mm, front 80mm.
IP class:
IP 41
The device is intended for indoor use.
Type B, according to the degree of protection against electric shock.
Important information
1 5