5
Chapter 3
Installation
Unpacking
Check to make sure that the package contains the following items:
#
Model 9020-OEM or Model 9020 CORRATER® transmitter
#
Six-foot
probe-to-instrument cable with Type B connector
#
P/N 710920 Test Probe (to verify 9020 instrument operation only)
#
User
Manual
Note:
All 9020-OEM and 9020 system components are carefully tested, inspected and
packaged prior to shipment. Before proceeding with the installation of the instrument, please
inspect all items for shipping damage and retain any damaged materials to support any claim
against the freight carrier should this become necessary.
Installation Procedure
Installation of the 9020-OEM and 9020 consists of three tasks:
#
Mechanical mounting of the transmitter
#
Electrical connections to the transmitter and probe
#
Verification of instrument operation
#
Selection of parameters for corrosion rate monitoring
Mechanical Mounting
The Model 9020-OEM is intended to me mounted in a customer supplied secondary
enclosure using four #10 machine screws through the mounting feet. Dimensions for the
mounting holes are shown in Figure 5.
Содержание CORRATER 9020
Страница 4: ...ii ...
Страница 16: ...CORRATER Transmitter User Manual 12 ...
Страница 18: ...CORRATER Transmitter User Manual 14 ...