obtained, then a shorter time cycle can be selected as long as the readings do not increase
substantially (>5%) over the readings taken with a 20 minute cycle.
Set the ALLOY MULTIPLIER switches to the appropriate multiplier value for the metal or alloy
of the electrodes as shown in the chart of Appendix 1. The switches represent the three
digits of the multiplier with the left-hand switch selecting the units value (0, 1 or 2); the center
switch selecting the 1/10's value (0.0-0.9); and the right-hand switch selecting the 1/100's
value 0.00-0.09).
Set the STANDARD/FLUSH toggle switch for STANDARD if using a probe with protruding
5.0 square centimeter electrodes or select FLUSH if using a probe with flush 0.5 square
centimeter electrodes.
Enable the power to the loop and after the initial measurement cycle has been completed, the
loop current will be continuously updated to the new value at the end of each measurement
cycle.