R&S ZVH
Cable and Antenna Test Mode
Configuring Cable and Antenna Tests
Operating Manual 1309.6946.12 - 09
49
If you do not have access to a PC, but still need a cable model that is not stored on the
R&S ZVH, you can also define the characteristics of a cable temporarily on the
R&S ZVH itself. It is not possible to save these in a data set, however. They are lost as
soon as you change them or load another cable model.
► Press the MEAS key.
► Press the "DTF Settings" softkey.
► Press the "Cable Config" softkey.
► Select the "Define User Model" menu item.
A submenu opens.
► Select the "Frequency" menu item.
► Enter the frequency of the cable you are testing.
► Select the "Velocity" menu item.
► Enter the cable velocity.
► Select the "Loss" menu item.
► Enter the cable loss.
You can now perform measurements on the temporary cable definition.
Note that as soon as you change any cable parameter or load another cable
model, the temporary data is lost and you have to define it again if you have to
perform further measurements.
After you have defined the cable characteristics, you still have to activate the use of the
customized cable.
► Press the "Cable Config" softkey.
► Select the "[ ] User Model" menu item.
If the R&S ZVH uses the customized cable model, it shows an [X] at the "User
Model" menu item.
2.2.1.3 Working with a DTF List
The DTF list shows the results of the distance to fault measurement in numerical form.
► Press the MEAS key.
► Press the "DTF Settings" softkey.
► Press the "DTF List" softkey.
The R&S ZVH opens a table that lists the peaks it has found during the
measurement.