R&S ZVH
Network Analyzer Mode (R&S ZVH-K42)
Performing Pulse Power Measurements (R&S ZVH-K29)
Operating Manual 1309.6946.12 - 09
177
Defining the output level of the tracking generator
The tracking generator generates a signal at the current R&S ZVH frequency. The
nominal output level of the signal is adjustable from 0 dBm to -40 dBm in steps of 1 dB.
► Press the SCALE/AMPT key.
► Press the "TG Power" softkey.
The R&S ZVH opens an input field to set the output level.
The output power of the tracking generator is in the range from -40 dBm and
0 dBm.
Tracking generator output level
Using an output level of 0 dBm (or a receiver attenuation of 0 dB) for S12 or S21
measurements can cause an overload at the RF input. It is therefore recommended to
use an output level of -10 dBm (default) or lower for these measurements.
This especially applies to measurements on amplifiers.