Preparing for Use
R&S
®
ZV-Z58/Z59
Mounting in a 19" Rack
Using the adapter ZZA-211 (order number 1096.3260.00) the unit can be mounted in 19" racks
according to the mounting instructions supplied with the rack adapter.
ATTENTION
Operation in a rack
•
Allow for sufficient air supply in the rack.
•
Make sure that there is sufficient space between the ventilation holes and the
rack casing.
EMI Protective Measures
In order to avoid electromagnetic interference, the unit may only be operated when it is closed and
with all shielding covers fitted. Only appropriate shielded signal and control cables may be used.
ATTENTION
Electrostatic discharge
To protect the calibration unit against Electrostatic Discharge (ESD) damage use the
wrist strap and grounding cord supplied with the network analyzer and connect
yourself to the GND connector at the front panel. For details refer to the Quick Start
uide of your analyzer.
G
Connecting the Unit to the AC Supply
The calibration unit is automatically adapted to the AC supply voltage supplied. The supply voltage
must be in the range 100 V to 240 V; 50 Hz to 60 Hz. The mains connector is located at the rear panel.
Connect the calibration unit to the AC power source using the AC power cable delivered with the
unit.
The maximum power consumption of the calibration unit is 125 W. The typical power consumption is
listed in the data sheet. The calibration unit is protected by two fuses which can be replaced as
described in section
Power on and off
The mains connector is located at the rear panel.
1164.0696.92 12