Command Reference
R&S
®
ZNB/ZNBT
883
User Manual 1173.9163.02 ─ 38
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC:EFE 'CH4TR1'
Display the generated trace in diagram area no. 2. No trace
number is assigned.
Usage:
Setting only
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED
<TraceName>
CALCulate<Ch>:PARameter:SDEFine
) to a diagram
area, using the
<WndTr>
suffix, and displays the trace. Use
WINDow<Wnd>]:TRACe<WndTr>:EFEed
to assign the trace to a diagram area with-
out using a numeric suffix.
Tip:
A trace can be assigned to a diagram only once. If a attempt is made to assign the
same trace a second time (e.g. by typing
DISP:WIND2:TRAC8:FEED 'CH4TR1'
after executing the program example below) an error message -114,"Header suffix out
of range" is generated. You can open the "Trace Manager" dialog to obtain an over-
view of all channels and traces, including the traces that are not displayed.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
).
<WndTr>
Trace number used to distinguish the traces of the same dia-
gram area
<Wnd>
.
Parameters:
<TraceName>
String parameter for the trace name, e.g. 'Trc4'.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:SHOW
<TraceName>[, <Boolean>]
Displays or hides an existing trace, identified by its trace name
<Trace_Name>
, or a
group of traces.
SCPI Command Reference