Rise Time Tab
The "Rise Time" tab allows to enable and configure the
Access:
SYSTEM – APPLIC > "TDR" > "Rise Time"
Access:
APPLIC key or "Application > Rise Time" menu
Rise Time
Enables/disables the rise time measurement.
The rise time measurement can only be performed if the active trace is
time domain transform is used. The latter,
in turn, requires the stimulus grid to be harmonic.
If the rise time measurement can not be enabled, a message is displayed:
Instead of performing the required configuration changes manually, you may also use
the "Auto Setup" button of the info field to let the firmware do this for you.
Remote command:
Start Value / Stop Value
Defines the lower/upper percentage for the rise time measurement.
Remote command:
CALCulate<Chn>:TTIMe:THReshold