GUI Reference
R&S
®
ZNB/ZNBT
494
User Manual 1173.9163.02 ─ 38
5.12.4 Time Domain Tab
Calculates and displays the results as a function of time.
Time domain analysis requires option R&S
ZNB-K2 / R&S
ZNBT-K2. If this option is
not installed, the "Time Domain X-Axis" tab is hidden.
Background information
Refer to
Chapter 4.7.2, "Time Domain Analysis"
For a comparison of the different transformation types and windows and for application
examples please also refer to the application note 1EZ44_OE which is available on the
Rohde & Schwarz internet at
http://www.rohde-schwarz.com/appnotes/1EZ44
5.12.4.1
Controls on the Time Domain Tab
"Low Pass Settings..." opens the
Time Domain
Selects the time domain representation for the active diagram area. The softkey is
enabled if a linear frequency sweep is active (see
on page 243). The ana-
lyzer automatically quits time domain representation as soon as a different sweep type
is selected.
The time domain results are obtained by transforming the measured frequency sweep
data into the time domain using an appropriate mathematical transformation type and
frequency window ("Impulse Response"). The sweep range and the output power for
the active channel is still displayed below the diagram; the displayed time interval is
shown in a second line:
Trace settings in time domain representation
While the time domain representation is active the trace settings behave as follows:
●
The "Start" and "Stop" settings in the "Time Gate" tab configure the time axis.
●
All trace formats including the circular diagrams are available.
Trace Config Softtool