GUI Reference
R&S
®
ZNB/ZNBT
401
User Manual 1173.9163.02 ─ 38
ous" sweep), where the measurement is continuously repeated without fixed time refer-
ence.
Any trigger event may start an entire sweep or a part of it. Moreover, it is possible to
switch off the RF source between consecutive triggered measurement sequences, and
to define a delay between trigger events and the measurement sequences.
The trigger settings are also valid for calibration sweeps. This means that, in external
trigger mode, the external trigger signal must be available during the system error cor-
rection, too. To start the calibration sweeps without delay, use the "FreeRun" trigger
type.
Background information
Refer to
Chapter 4.1.4.1, "Partial Measurements and Driving Mode"
Output trigger
The R&S
ZNB/ZNBT provides a configurable output trigger signal to synchronize exter-
nal devices with the measurement. This signal is available at the rear panel connector
EXT TRIG OUT. Configuration of the output trigger signal is a remote control feature
(
TRIGger:CHANnel<Ch>:AUXiliary...
5.6.1.1
Controls on the Trigger Tab
FreeRun / External / Manual / Multiple Triggers
The four buttons select the source of the trigger event:
●
In "FreeRun" mode a new measurement is started immediately without waiting for
a trigger event and without fixed time reference. The remaining trigger settings are
not valid.
Trigger Softtool