GUI Reference
R&S
®
ZNB/ZNBT
271
User Manual 1173.9163.02 ─ 38
Delete
Deletes the selected calibration from the list of configured calibrations.
Modify...
Edits the selected system error correction: opens the
with the
corresponding ports and calibration type pre-selected.
Next
.
Inactive as long as no calibration is defined.
Define Calibration Dialog
In MultiCal mode, the "Define Calibration" dialog allows to define the individual system
error corrections by selecting the test ports to be calibrated and the calibration unit,
characterization and calibration type to be used.
Access:
CHANNEL – CAL > "Start Cal" > "Calibration" – "Start... (Cal Unit)" >
"Add..." / "Modify..."
●
The "Multiple Cal in Calibration Wizard" (MultiCal) feature is disabled by default
and has to be activated in the
tab of the "System Config" dialog.
●
For background information on calibration types see
The layout of the main panel depends on the number of test ports available.
Figure 5-7: Define Calibration dialog (MultiCal, > 4 test ports)
Ports
Selects the test port(s) to be calibrated.
Note:
Calibration and port de-/activation.
Cal Softtool