Calibration Type / Source
Selects the calibration type for the selected physical port(s). For an overview refer to
.
The reflection calibration types can be used for any combination of physical ports:
Reflection calibrations are simply repeated for each selected port. A transmission cali-
bration type requires at least two physical ports. For the
unidirectional
transmission
calibration types ("Trans Norm", "One Path Two Ports"), the direction (
"Source"
port)
must be specified in addition.
Note:
The transmission normalizations and the "One Path Two Ports" calibration type
require two-port (Through) characterization data for the cal unit, which may not be
available on older calibration units or in user-defined cal unit characterization files. If a
tooltip indicates missing two-port characterization data, simply perform a new charac-
terization of your cal unit. In the first dialog of the "Characterization" wizard, select
"Take All OSM and Through" to make sure the necessary two-port data is acquired.
See also
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Step 2: Connections
Defines the port assignment(s) between the R&S
ZNB/ZNBT and the calibration unit.
Background information
Refer to
Chapter 4.5.5.4, "Multiple Port Assignments"
on page 164 for details on multi-
ple port assignments
Access:
CHANNEL > CAL > Start Cal > Start... (Cal Unit) > Next
Cal Softtool