GUI Reference
R&S
®
ZNB/ZNBT
329
User Manual 1173.9163.02 ─ 38
Characterization Data
Displays all characterizations which are stored on the active cal unit. The "Factory"
characterization is available for all calibration units; it ensures an accurate calibration
for all standard applications. Characterizations stored on an SD card (inserted at the
cal unit) are prefixed with "SD:".
Tip:
Characterizations stored on an SD card (inserted at the cal unit) are prefixed with
"SD:".
The properties of the selected characterization are shown below the list. "Delete" dele-
tes the selected characterization file; "Start Characterization..." opens the
to create a new characterization.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Characterization Wizard
The "Characterization" wizard guides you through the automatic characterization of a
calibration unit.
Access:
The guided characterization consists of the following steps:
1.
"Characterization":
Select the characterized ports and cal unit standards and ini-
tiate the characterization sweeps.
2.
"Save Characterization Data":
Save the characterization data to the calibration
unit.
Step 1: Characterization
Selects the calibration type as well as the characterized cal unit ports and initiates the
necessary characterization sweeps.
Characterization procedure
To acquire accurate characterization data, the test setup must be properly calibrated
before you start the characterization wizard. Use the calibration type that you wish to
perform with your new cal unit characterization; see
Chapter 4.5.5.3, "Characterization
Cal Softtool