Command Reference
R&S
®
ZNB/ZNBT
1169
User Manual 1173.9163.02 ─ 38
Manual operation:
See
SYSTem:PRESet[:DUMMy]
Performs a factory preset of all instrument settings (i.e. all open recall sets) or of the
active recall set, depending on the
SYSTem:PRESet:SCOPe
settings. The command is
equivalent to
*RST
and to the action of the PRESET key on the front panel.
Note:
If a user-defined preset is active (
ON
), the
PRESET
key initiates a user-defined preset, while
SYSTem:PRESet[:DUMMy]
and
*RST
still activate the factory preset. If "Align *RST to User Defined Preset" is selected
in the "System Configuration > Presets" dialog and a valid user preset file is available,
*RST
and
SYSTem:PRESet[:DUMMy]
also restore the user-defined settings.
Example:
SYST:PRES:SCOP SING
Define the scope of a preset: the active recall set is reset only.
SYST:PRES
Reset the parameters of the current recall set.
Manual operation:
See
SYSTem:PRESet:REMote[:STATe]
<arg0>
Defines the behavior of the
*RST
and
●
Off (default):
*RST
and
SYSTem:PRESet
restore the factory preset settings.
●
On:
If a valid user preset file is available,
*RST
and
SYSTem:PRESet
restore the
user-defined settings.
Parameters:
<arg0>
Manual operation:
See
SYSTem:PRESet:SCOPe
<Scope>
Specifies whether a preset (
SYSTem:PRESet[:DUMMy]
;
*RST
) affects the active
recall set only or all open recall sets.
Parameters:
<Scope>
ALL | SINGle
ALL
–
all open recall sets are deleted and the recall set "Set1" is
created with default trace and channel settings.
SINGle
–
the settings of the active setup are reset; the name of
the active setup and the parameters of all other setups remain
unchanged.
Example:
See
Manual operation:
See
SCPI Command Reference