Command Reference
R&S
®
ZNB/ZNBT
1021
User Manual 1173.9163.02 ─ 38
Note:
The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
is enabled
only after steps 1 and 2 have been carried out (
CORRection:IMODulation:LTONe[:ACQuire]
CORRection:IMODulation:UTONe[:ACQuire]
).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
).
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNB/ZNBT-K4).
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation:
See
SCPI Command Reference