Troubleshooting and Repair
R&S
®
TS7124AS
59
Instructions Handbook 1179.2948.02 ─ 01
10
Troubleshooting and Repair
Every
is allowed to perform the activities described in this
chapter. Any
repair
activities are allowed for Rohde
&
Schwarz
Chapter 5, "Transportation, Handling and Storage"
10.1
Door Error
When a door error occurs, the control system automatically deactivates the chamber
by depressurizing the pneumatic system.
You can verify a door error by sending a
ERR
", the door
has an error.
To resolve this error, proceed as follows:
1. Disconnect the chamber from its electric power supply.
2. If the door is blocked, which prevents it from closing within the
remove any obstacle that blocks the door.
Chapter 7.1, "Activating the Chamber"
4. If the
period is shorter than the time that it takes the door to open or
close, set the
TIMEOUT
value to match the door speed. See
Alternatively, adjust the door speed to match the
TIMEOUT
value.
5. If the door is pushed to close while the
select one of the following:
state.
● Leave it disabled, but make sure that no operator pushes the door to close.
6. If there is a controller conflict, resolve any issues described in
7. If the door still has an error and the "
ERR
" reply persists, inform Rohde
&
Schwarz
customer support.
10.2
Controller Conflicts
You can control the chamber by any of the following devices:
●
"
IRC
": Internal remote control module (labeled 1 in
●
"
MAN
●
"
BOTH
": IRC + MAN at the same time
Controller Conflicts