Inspection and Maintenance
Familiarize yourself with residual risks and potentially dangerous situations.
See
Chapter 2.3, "Potentially Danger-
The chamber has no factory default settings, except the
......................................................................................... 55
..................................................................................... 55
Preparing the Chamber for Maintenance
................................................................56
...............................................................................56
9.1
Recommended Intervals
To ensure safe operation and to retain the functional readiness and long operational
life of the chamber, perform the inspection and maintenance tasks as scheduled:
Table 9-1: Scheduled inspection and maintenance
Maintenance interval
Maintenance tasks
Daily
Chapter 9.4.1, "Daily Functional Check"
Weekly
Chapter 9.4.2, "Checking the Absorber"
Every 100
000 cycles
Chapter 9.4.4, "Cleaning the Gasket"
If necessary
Chapter 9.4.3, "Cleaning the Chamber"
Every time you calibrate test instruments
Chapter 9.4.5, "System Calibration"
Once a year
The intervals in
are recommended for 160 operating hours per month. If you
operate the chamber for more hours, adapt the maintenance intervals accordingly.
9.2
Regular Safety Inspections
Daily safety check
► Before operation, test the door's automatic emergency deactivation mechanism for
proper functioning. See
Chapter 6.6, "Testing Safety Systems"