Installation and Commissioning
R&S
®
TS7124AS
35
Instructions Handbook 1179.2948.02 ─ 01
a) Remove the safety cap (
) from the control connector [X21] on the
rear side of the chamber.
Figure 6-10: Safety cap, covering the D-Sub 25-pin connector [X21]
b) Keep the safety cap for future use.
c) Connect the push-button switch unit to connector [X21].
d) Position the push-button switch unit in a place that avoids its collision with the
chamber's opening door.
The operator must be able to reach the control unit easily without entering the
door zone (see
2. For remote control via a
serial port
, connect an RS-232 cable from your control
system to the D-Sub
9 connector (see
[X26]
For port settings, see
Use a "straight-through" RS-232 cable, see
"Use a straight-through cable for the
RS-232 connection to the control computer"
reasons, limit the length of the RS-232 cable to a maximum of 30
m.
3. For remote control via
Ethernet
, connect a LAN cable from your control system to
the RJ45 connector (see
[X20]
in
).
You can use a LAN cable of arbitrary length. For Ethernet configuration, see
If you want to change the interface, proceed as described in
Connecting to Control and Power