Remote control commands
R&S
®
SMM100A
995
User Manual 1179.1341.02 ─ 05
●
Under
TRIGger
several sweep subsystems exist.
●
The trigger source names correspond directly to the various settings of manual
control. SCPI uses different names which are also accepted by the instrument. The
SCPI names should be used if compatibility is an important consideration.
In addition to these commands, see more trigger-related commands in the modulation
and RF signal subsystems.
Table 13-5: Cross-reference between the manual and remote control
R&S proprietary value name
SCPI conform value name
Parameter in manual control
AUTO
IMMediate
"Auto" mode
SINGle
BUS
"Single" mode.
EXTernal
EXTernal
"Ext Single" and "Ext Step" mode.
Use command
LFO:SWEep:MODE
to select between the two sweep
modes.
EAUTo
-
"Ext Start/Stop" mode.
....................................................................................995
.................................................................................995
....................................................................................995
....................................................................................995
:TRIGger<hw>:FSWeep[:IMMediate]
................................................................................996
:TRIGger<hw>:LFFSweep:IMMediate
.............................................................................. 996
:TRIGger<hw>:PSWeep[:IMMediate]
................................................................................996
:TRIGger<hw>[:SWEep][:IMMediate]
................................................................................996
.............................................................................................. 997
:TRIGger<hw>:FSWeep:SOURce
<Source>
:TRIGger<hw>:LFFSweep:SOURce
<Source>
:TRIGger<hw>:PSWeep:SOURce
<Source>
:TRIGger<hw>[:SWEep]:SOURce
<Source>
Selects the trigger source for the corresponding sweeps:
●
FSWeep
- RF frequency
●
LFFSweep
- LF frequency
●
PSWeep
- RF level
●
SWEep
- all sweeps
The source names of the parameters correspond to the values provided in manual
control of the instrument. They differ from the SCPI-compliant names, but the instru-
ment accepts both variants.
Use the SCPI name, if compatibility is an important issue. Find the corresponding
SCPI-compliant commands in
Cross-reference between the manual and remote con-
Setting parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
TRIGger subsystem