Remote control commands
R&S
®
SMM100A
971
User Manual 1179.1341.02 ─ 05
Parameters:
<MmemProtState>
0 | 1 | OFF | ON
*RST:
0
Setting parameters:
<SecPassWord>
string
Current security password
The default password is 123456.
Example:
SYSTem:SECurity:VOLMode:STATe "123456", 1
SYSTem:REBoot
Manual operation:
See
:SYSTem:COMMunicate:GPIB:LTERminator
<LTerminator>
Sets the terminator recognition for remote control via GPIB interface.
Parameters:
<LTerminator>
STANdard | EOI
EOI
Recognizes an LF (Line Feed) as the terminator only when it is
sent with the line message EOI (End of Line). This setting is rec-
ommended particularly for binary block transmissions, as binary
blocks may coincidentally contain a characater with value LF
(Line Feed), although it is not determined as a terminator.
STANdard
Recognizes an LF (Line Feed) as the terminator regardless of
whether it is sent with or without EOI.
*RST:
n.a. (factory preset: STANdard)
Example:
See
Example "Finding out the used VISA resource strings"
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command
.
Return values:
<Resource>
string
Example:
See
Example "Finding out the used VISA resource strings"
Usage:
Query only
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
Sets the GPIB address.
SYSTem subsystem