Remote control commands
R&S
®
SMM100A
751
User Manual 1179.1341.02 ─ 05
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <AM1>; LF2 <AM2>
Example:
See
Example "Creating an amplitude modulated RF signal"
Manual operation:
See
[:SOURce<hw>]:AM<ch>[:DEPTh]
<Depth>
Sets the depth of the amplitude modulation in percent.
Suffix:
<ch>
.
1..2
Modulation signal channel
Parameters:
<Depth>
float
Range:
0 to 100
Example:
See
Example "Creating an amplitude modulated RF signal"
Manual operation:
See
[:SOURce<hw>]:AM:DEPTh:SUM
<AmDepthSum>
Sets the total depth of the LF signal when using combined signal sources in amplitude
modulation.
Parameters:
<AmDepthSum>
float
Range:
0 to 100
Increment: 0.1
*RST:
30
Example:
See
Example "Using combined LF signals"
Manual operation:
See
[:SOURce<hw>]:AM:DEViation:MODE
<AmDevMode>
Selects the coupling mode. The coupling mode parameter also determines the mode
for fixing the total depth.
Parameters:
<AmDevMode>
UNCoupled | TOTal | RATio