Configuring the RF signal
R&S
®
SMM100A
393
User Manual 1179.1341.02 ─ 05
"Fix"
Fixes the last set sensitivity setting.
Remote command:
[:SOURce<hw>]:POWer:ALC:DSENsitivity
Driver Amplifier
Selects the switching state of the driver amplifier, that means the switchable amplifier
or step attenuator at the output.
"Auto"
Switches the step attenuator automatically.
"On/Off"
The driver amplifier is switched on, or off respectively.
"On (maximum Gain)"
The driver amplifier is switched on.
The R&S
SMM100A supplies the maximum level at the output within
the frequency range 6
GHz to 10
GHz.
Note:
In this mode, the harmonics increase.
"Fix"
Fixes the last setting.
The
button switches the driver amplifier state first to auto-
matic, calculates all level settings and then resets to fix.
Remote command:
[:SOURce<hw>]:POWer:ALC:DAMPlifier
Sample Level
Determines the level at the output of the instrument during the sample process.
How to: See
Chapter 7.12.2.2, "How to enable the ALC"
"Full"
Attenuates the sample level by the crest factor of the used I/Q modu-
lation signal or 9
dB, whichever is lower. Additionally, an internal
headroom of the signal is considered.
"Attenuated"
Sets an additional attenuation of 40
dB.
"Minimum"
Sets the attenuation to maximum.
Remote command:
[:SOURce<hw>]:POWer:ALC:SLEVel
7.12.2.2
How to enable the ALC
1. Select "RF" > "RF Level" > "Automatic Level Control".
2.
NOTICE!
Risk of DUT damage due to high input power. The ALC modes "State" >
"Off (Sample & Hold)" and "State" > "Sample & On" use a short sample pulse to set
the level control. With the setting "Sample Level" > "Full", this sample pulse results
in a CW RF pulse. Usually, the RMS value of this CW pulse equals the configured
RMS level value of the signal, but it can also rise up to the PEP level.
To protect the DUT from damage due to high input power, use the following setting
combinations for the sampling states "Off (Sample & Hold)" or "Sample & On":
● Set "Sample Level " > "Attenuated" or
● Set "Sample Level " > "Minimum".
Improving level performance