Remote control commands
R&S
®
SMM100A
935
User Manual 1179.1341.02 ─ 05
Example:
POW:ALC:MODE?
Response:
ONTable
"Table and On" is set automatically.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Activates automatic level control in the selected mode.
How to: See
Chapter 7.12.2.2, "How to enable the ALC"
Parameters:
<State>
AUTO | OFFTable | ON | ONSample | ONTable | OFF
AUTO
Adjusts the output level to the operating conditions automati-
cally.
OFFTable
Controls the level with the attenuation values of the internal ALC
table.
ON
Activates internal level control permanently.
OFF
Deactivates internal level control, "Sample & Hold" mode is
active.
ONSample
Starts the internal level control with the first change.
ONTable
Starts with the attenuation setting from the table and continues
with automatic level control.
*RST:
AUTO
For more details on the individual settings, an overview of the
functionality and details on what is to be considered, see
Example:
POW:ALC ON
Activates internal level control.
Manual operation:
See
[:SOURce<hw>]:POWer:ALC:DAMPlifier
<Amplifier>
Selects the driver amplifier switching state.
Parameters:
<Amplifier>
OFF | ON | AUTO | FIX | ONMG
AUTO
Switches the attenuator automatically.
SOURce subsystem