Configuring the internal baseband source
R&S
®
SMM100A
212
User Manual 1179.1341.02 ─ 05
4.9.3.3
Level / clock / marker settings
Access:
1. Select "Baseband > ARB > General > Multi Segment".
2. In the "ARB: Multi Segment > General" dialog, select an existing list.
3. Select "ARB: Multi Segment > Segment Table" and configure the segments.
4. Select "ARB: Multi Segment > Level/Marker/Clock".
The "Level/Marker/Clock" tab provides the parameters necessary to adjust the
level, marker and clock settings for the selected multi-segment waveform file.
Settings:
Level
Defines the way the instrument determines the output level of each of the multiple
waveforms.
"Unchanged"
Concerning the level settings, the segments are output exactly as
defined in the individual files. The "Level" display applies only to the
segment with the highest RMS value. Sometimes, the remaining seg-
ments are output at a lower level than the displayed value.
"Equal RMS"
Segments are output so that all segments have the same RMS value.
The "Level" display applies to all segments.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:LEVel[:MODE]
Clock
Defines the way the instrument determines the clock rate of each of the multiple wave-
forms.
Generating multi-segment waveform files