Remote Control Commands
R&S
®
SMC100A
210
Operating Manual 1411.4060.32 ─ 04
●
Activating the original state of delivery (factory reset,
on page 210). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:SOURce<hw>:PRESet
The command presets all parameters which are related to the selected signal path.
Fading (if available) and transient recorder are only preset by command
*RST
.
Example:
SOUR:PRES
presets all settings that are related to signal path
SYSTem:PRESet
The command triggers an instrument reset. It has the same effect as:
●
the PRESET key on the front panel
●
the
*RST
command
For an overview of the settings affected by the preset function, see
Instrument Settings - Preset Key"
Example:
SYST:PRES
All instrument settings (also those that are not currently active) are
reset to their default values.
Usage:
Setting only
:SYSTem:FPReset
The command triggers an instrument reset to the original state of delivery.
Note:
Since Factory Preset resets the Remote Channel and network settings to the
default values, executing Factory Preset via remote control terminates the connection to
the instrument, if these settings had been configured to values different to the default
ones!
The factory preset function resets nearly all instrument settings. In addition to the regular
preset by means of the PRESET key, a "Factory Preset" resets also the following values:
●
Reference frequency settings ("Ref Oscillator" menu)
●
Power on settings ("Level/EMF" menu)
●
Network settings including hostname ("Setup" menu)
●
Remote Channel settings including GPIB address ("Setup" menu)
●
Start/Stop Gui Update ("Setup" menu)
●
Display and keyboard settings ("Setup" menu).
To maintain security, password settings and all settings protected by these passwords
like disabled USB and LAN connections are not changed.
Not affected by the "Factory Preset" are also user data, lists or instrument settings files,
created for example by means of the Save/Recall function.