Remote Control
R&S
®
RTM20xx
358
User Manual 1317.4726.02 ─ 01
Parameters:
<EventMode>
OFF | EACH | SINGle | CYCLic
OFF
No action is executed.
EACH
The selected action is executed on each violation of the mask.
SINGle
The selected action is executed once after the n-th violation.
CYCLic
The selected action is executed repeatedly after each n-th viola-
tion.
The number of violations <n> is set with the relevant
*RST:
OFF
Firmware/Software:
FW 03.800
MASK:ACTion:SOUNd:EVENt:COUNt
<EventCount>
MASK:ACTion:STOP:EVENt:COUNt
<EventCount>
MASK:ACTion:SCRSave:EVENt:COUNt
<EventCount>
MASK:ACTion:PRINt:EVENt:COUNt
<EventCount>
MASK:ACTion:WFMSave:EVENt:COUNt
<EventCount>
Sets the number of mask violations after which the action is executed. The command is
only relevant if the associated
SINGle
or
CYCLic
.
Parameters:
<EventCount>
Integer value, number of mask violations
Firmware/Software:
FW 03.800
15.2.12 Search
...............................................................................359
...................................................................................362
..................................................................................362
...................................................................................364
Rise/Fall Time Search Configuration
....................................................................365
....................................................................................366
Data2Clock Search Configuration
.........................................................................368
................................................................................370
Command Reference