Remote Control
R&S
®
RTM20xx
370
User Manual 1317.4726.02 ─ 01
Firmware/Software:
FW 03.800
SEARch:TRIGger:DATatoclock:STIMe
<SetupTime>
Sets the minimum time
before
the clock edge while the data signal must stay steady
above or below the data level. The setup time can be negative. In this case, the setup
interval starts after the clock edge, and the hold time must be positive and longer than
the absolute value of the setup time.
Parameters:
<SetupTime>
Range:
depends on time base and sample interval
Firmware/Software:
FW 03.800
15.2.12.8
Pattern Search Configuration
...............................................................................370
..............................................................................370
..............................................................................371
...................................................................371
......................................................................371
.....................................................................372
.......................................................................372
SEARch:TRIGger:PATTern:SOURce
<Pattern>
Specifies the search pattern - the state for each channel.
Parameters:
<Pattern>
String parameter
String containing 0, 1, X|x for each channel. The order of channels
is fixed: CH1 CH2 [CH3 CH4].
Example:
SEAR:TRIG:PATT:SOUR '1X10'
CH1, CH3 are high, CH4 is low. These states are logically com-
bined with
not matter (don't care).
SEARch:TRIGger:PATTern:FUNCtion
<Function>
Sets the logical combination of the channel states.
Command Reference