Putting into Operation
R&S
®
RT-ZZ80
17
User Manual 1409.7689.02 ─ 02
screen. If the DUT is not floating but ground-referenced, an auto zero routine may
improve the measurement results.
1. Connect the probe to the R&S
RTO oscilloscope and configure the probe as
described in
chapter 2.1, "Connecting the Probe to the R&S RTO Oscillo-
2. Short the signal pin and the ground pin together and connect them to the
ground of the DUT.
3. Start the auto zero procedure on the R&S
RTO:
"Vertical" menu > "Probe Setup" > "Ch" tab > "Auto zero"
Auto Zero