![Rohde & Schwarz RT-ZD40 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/rohde-and-schwarz/rt-zd40/rt-zd40_user-manual_1477678047.webp)
Functional Check
R&S
®
RT-ZD40
47
User Manual 1410.5363.02 ─ 05
7
Functional Check
The functional check confirms the basic operation of the R&S
RT-ZD40 differen-
tial probe. The functional check is not suitable for verifying compliance with the
probe specifications.
1. Connect the R&S
RT-ZD40 to a Rohde
&
Schwarz oscilloscope as described
in
Chapter 2.1, "Connecting the Probe to the Oscilloscope"
2. Connect a socket adapter to the signal sockets.
3. Using a short lead and a mini clip, connect one of the signal sockets to the
square wave output of the oscilloscope.
4. Using a short lead and a mini clip, connect the other signal socket to the
probe ground connector of the oscilloscope.
5. Press the [Preset] key and then the [Autoset] key on the oscilloscope.
A square wave with 1
V amplitude between 0
V and 1
V is displayed on the
display.
6. Reverse the pins at the probe and repeat
.