Remote Control Commands
R&S
®
RTM3000
634
User Manual 1335.9090.02 ─ 08
16.11.5.2
CAN - Trigger
To configure the protocol trigger, make sure to set first:
●
BUS
●
SBUS1 | SBUS2
............................................................................................ 635
.............................................................................................. 635
...........................................................................................636
...............................................................................................636
............................................................................................. 637
.............................................................................................. 637
............................................................................................637
TRIGger:A:CAN:TYPE
<TriggerType>
Specifies the trigger mode for CAN.
Parameters:
<TriggerType>
STOFrame | EOFrame | ID | IDDT | FTYPe | ERRCondition
STOFrame
Start of frame
EOFrame
End of frame
ID
Sets the trigger to a specific message "Identifier" or an "Identi-
fier" range.
Specify the identifier with
.
IDDT
Sets the trigger to a combination of "Identifier and Data" condi-
tion. The instrument triggers at the end of the last byte of the
specified data pattern.
Specify the "Identifier" (see
ID
), and the "Data" with
and
FTYPe
Triggers on a specified "Frame".
Specify the frame type with
.
Serial Bus Analysis