TRIGger:A:MODE
<TriggerMode>
Sets the trigger mode. The trigger mode determines the behavior of the instrument if
no trigger occurs.
Parameters:
<TriggerMode>
AUTO | NORMal
AUTO
The instrument triggers repeatedly after a time interval if the trig-
ger conditions are not fulfilled. If a real trigger occurs, it takes
precedence.
NORMal
The instrument acquires a waveform only if a trigger occurs.
*RST:
AUTO
TRIGger:A:SOURce
<Source>
Sets the trigger source for the selected A trigger type.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | EXTernanalog | LINE | SBUS1 |
SBUS2 | D0..D15
CH1 | CH2 | CH3 | CH4
One of the analog input channels is the trigger source. Available
channels depend on the instrument type.
EXTernanalog
External trigger input on the front panel
LINE
AC power supply line for the line trigger
SBUS1 | SBUS2
Serial bus 1 or 2
Requires at least one protocol option for serial bus (R&S
RTM-
K1 to K3)
D0..D15
Digital channels D0 to D15, can be used as trigger sources for
edge, width, timeout and pattern trigger. Require MSO option
R&S
RTM-B1.