3.4
Programming
The following sections provide programming examples for the NB-IoT multi-evaluation
measurement.
.................................................................................................248
.............................................................................................. 252
3.4.1
General Examples
The NB-IoT multi-evaluation measurement is programmed as follows:
●
The measurement is controlled by SCPI commands with the following syn-
tax:
...NIOT:MEAS:MEValuation...
●
Use general commands of the type
...:NIOT:MEAS...
(no
:MEValuation
mnemonic) to define the signal routing and perform RF and analyzer settings.
●
After a
*RST
, the measurement is switched off. Use
READ:NIOT:MEAS:MEValuation...?
to initiate a single-shot measurement and
retrieve the results. You can also start the measurement using
INIT:NIOT:MEAS:MEValuation
and retrieve the results using
FETCh:NIOT:MEAS:MEValuation...?
.
●
For synchronization and proper decoding, some UE signal settings must be in
accordance with the measured signal; see
Chapter 3.4.1.2, "Specifying Required
3.4.1.1
Specifying General and Common Measurement Settings
// *****************************************************************************
// Initial system-reset
// *****************************************************************************
*RST; *OPC?
*CLS; *OPC?
// *****************************************************************************
// Define signal routing and perform RF and analyzer settings:
// Center frequency 1850 MHz, frequency offset 1 kHz, peak power -2 dBm,
// no user margin and 1 dB mixer level offset.
// *****************************************************************************
ROUTe:NIOT:MEAS:SCENario:SALone RF1C, RX1
CONFigure:NIOT:MEAS:RFSettings:EATTenuation 2
CONFigure:NIOT:MEAS:BAND OB2
CONFigure:NIOT:MEAS:RFSettings:FREQuency 1850E+6
CONFigure:NIOT:MEAS:RFSettings:FOFFset 1000
CONFigure:NIOT:MEAS:RFSettings:ENPower -2
CONFigure:NIOT:MEAS:RFSettings:UMARgin 0
CONFigure:NIOT:MEAS:RFSettings:MLOFfset 1
Programming