NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
249
User Manual 1178.5689.02 ─ 06
3.4.1.2
Specifying Required Settings
// *****************************************************************************
// Specify required UE signal settings:
// subcarrier spacing, NPUSCH format, resource allocation, physical cell ID,
// delta sequence shift value, frequency sync range, modulation scheme.
// *****************************************************************************
CONFigure:NIOT:MEAS:MEValuation:SCSPacing S15K
CONFigure:NIOT:MEAS:MEValuation:NPFormat F1
CONFigure:NIOT:MEAS:MEValuation:SUBCarrier 3, 6
CONFigure:NIOT:MEAS:MEValuation:NRUNits NRU04
CONFigure:NIOT:MEAS:MEValuation:NREPetitions NR2
CONFigure:NIOT:MEAS:MEValuation:PLCid 10
CONFigure:NIOT:MEAS:MEValuation:DSS 5
CONFigure:NIOT:MEAS:MEValuation:FSYRange HIGH
CONFigure:NIOT:MEAS:MEValuation:MODulation:MSCHeme QPSK
3.4.1.3
Specifying Measurement-Specific Settings
// *****************************************************************************
// Define stop condition (stop on limit failure), error handling, timeout and
// number of slots per captured sequence.
// *****************************************************************************
CONFigure:NIOT:MEAS:MEValuation:SCONdition SLFail
CONFigure:NIOT:MEAS:MEValuation:MOEXception ON
CONFigure:NIOT:MEAS:MEValuation:TOUT 3600
CONFigure:NIOT:MEAS:MEValuation:NSLots 20
// *****************************************************************************
// Specify modulation measurement settings: 20 statistics cycles, no exclusion
// *****************************************************************************
CONFigure:NIOT:MEAS:MEValuation:SCOunt:MODulation 20
CONFigure:NIOT:MEAS:MEValuation:MODulation:EEPeriods:NPUSch:LEADing OFF
CONFigure:NIOT:MEAS:MEValuation:MODulation:EEPeriods:NPUSch:LAGGing OFF
// *****************************************************************************
// Specify spectrum measurement settings:
// measurement length of 30 slots, for ACLR measure all adjacent channels.
// *****************************************************************************
CONFigure:NIOT:MEAS:MEValuation:SCOunt:SPECtrum:ACLR 30
CONFigure:NIOT:MEAS:MEValuation:SCOunt:SPECtrum:SEMask 30
CONFigure:NIOT:MEAS:MEValuation:SPECtrum:ACLR:ENABle ON, ON
// *****************************************************************************
// Specify power measurement settings: 30 statistics cycles
// *****************************************************************************
CONFigure:NIOT:MEAS:MEValuation:SCOunt:POWer 30
Programming