Performing measurements
R&S
®
NRPM
73
Mannual 1425.8663.02 ─ 08
Duty Cycle
Sets the duty cycle, the percentage of one period during which the signal is
active, for pulse modulated signals. If activated, the sensor calculates the signal
pulse power from its value and the average power.
Remote command:
[SENSe<Sensor>:]CORRection:DCYCle:STATe
[SENSe<Sensor>:]CORRection:DCYCle
Smoothing
Activates the smoothing filter, a steep-cut off digital lowpass filter. The filter
reduces result fluctuations caused by modulation.
Remote command:
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
Trace mode
Provides the parameters of the trace measurement.
Access:
► In the measurement navigation pane of the web browser-based user interface
main dialog, select "Trace".
The remote commands required to define the trace settings are described in
Chapter 10.7.3, "Configuring a trace measurement"
................................................................................................. 74
Trace Time
Sets the trace length.
Remote command:
Measurement applications